Noise levels in a dental teaching institute: a matter of concern!
Mostra el registre complet de l'element
Visualització
(383.1Kb)
|
|
|
|
|
|
Singh, Simarpreet; Gambhir, Ramandeep Singh; Singh, Gurminder; Sharma, Sumit; Kaur, Amarinder
|
|
Aquest document és un/a article, creat/da en: 2012
|
|
|
|
Objective: To measure and assess the noise levels produced by various dental equipments in different areas of a
dental institution and to recommend improvements if noise levels are not within permissible limits.
Materials and methods: Sound levels were measured at three different areas of a dental institution where learning
and teaching activities are organized. The sound level was measured using a sound level meter known as 'decibulolmeter'.
In each area the noise level was assessed at two positions-one, at 6 inches from the operators ear and
second, at the chairside instrument trolley. Noise levels were also assessed from a central location of the clinic area
when multiple equipments were in operation simultaneously.
Results: Dental laboratory machine, dental hand-piece, ultrasonic scalers, amalgamators, high speed evacuation,
and other items produce noise at different sound levels which is appreciable. The noise levels generated varied
between 72.6 dB in pre-clinics and 87.2 dB in prosthesis laboratory. The results are comparable to the results of
other studies which are conducted elsewhere. Although the risk to the dentists is lesser, but damage to the hearing
is possible over prolonged periods.
Conclusion: Higher noise levels are potentially hazardous to the persons working in such environments especially
in the laboratory areas where noise levels are exceeding the permissible limits.
|
|