Comparative study of mandibular linear measurements obtained by cone beam computed tomography and digital calipers
Mostra el registre complet de l'element
Visualització
(417.9Kb)
|
|
|
|
|
|
Tarazona Álvarez, Pablo; Romero Millán, Javier; Peñarrocha Oltra, David; Fuster Torres, Mª Ángeles; Tarazona Álvarez, Beatriz; Peñarrocha Diago, Miguel
|
|
Aquest document és un/a article, creat/da en: 2014
|
|
|
|
Objectives: Cone beam computed tomography (CBCT) is an innovative dental of imaging system characterized by
rapid volumetric imaging with patient exposure to a single dose of radiation. The present study was carried out to
compare the linear measurements obtained with CBCT and digital caliper in 20 mandibles from human cadavers.
Study design: A total of 4800 linear measurements were measured between different mandibular anatomical po
-
ints with CBCT and digital caliper. The real measurements were defined as those obtained with the digital caliper.
Posteriorly, the mandibles were scanned to obtain the CBCT images, with software-based measurements of the
distances.
Results: The measurements obtained with the digital caliper were greater. The CBCT technique underestimated
distances greater than 100 mm.
Conclusions: CBCT allows to obtain linear mandibular anatomical measurements equivalent to those obtained with
digital caliper. The differences existing between both methods were clinically acceptable.
|
|
Veure al catàleg Trobes
|
Aquest element apareix en la col·lecció o col·leccions següent(s)