Pseudotumor tras cirugía de revisión por rotura de cabeza cerámica en una artroplastia total de cadera con par cerámica-cerámica: caso clínico
Mostra el registre complet de l'element
Visualització
(423.4Kb)
|
|
|
|
|
|
Sangüesa, M. J.; Cabanes, F.; Fernández Gabarda, Rafael; Valverde, C.
|
|
Aquest document és un/a article, creat/da en: 2011
|
|
|
|
We report a case presenting an important and early metallosis. Three months after a total hip arthroplasty
primary using ceramic-ceramic pair, the patient sustained ceramic head fracture was revised to a metal-on-polyethylene
articulation. Fourteen months later, a metallosis pseudotumor caused by the fracture of the metal head was
observed. The patient was then revised to a metal-metal pair preserving the femoral component since it was well
anchored while the Morse cone connection had a small impact. Two years after the last surgery the results are satisfactory.Se presenta un caso clínico con una importante y precoz metalosis, tan solo tres meses después de la
revisión de una artroplastia total de cadera primaria par cerámica-cerámica, en la que se había roto la cabeza cerámica,
y en la revisión se sustituyó por un par metal-polietileno. Catorce meses después apareció un pseudotumor causado
por la metalosis por rotura de la cabeza metálica, por lo que se realizó una nueva revisión utilizando un par
metal-metal preservando el vástago femoral que estaba perfectamente anclado aunque el cono Morse tenía una pequeña
agresión. Dos años después de esta última cirugía los resultados son satisfactorios. |
|