Evaluation of primary stability of innovated orthodontic miniscrew system (STS): an ex-vivo study
Mostra el registre complet de l'element
Visualització
(1.318Mb)
|
|
|
|
|
|
Seifi, Massoud; Matini, Negin-Sadat
|
|
Aquest document és un/a article, creat/da en: 2016
|
|
|
|
Background: Stability is determined as one of the requirements in use of Temporary Anchorage Devices (TAD) in
orthodontics. Miniscrew has been a widely used Bone Anchor. Compared with mini-implant that necessitates osseointegration;
mechanical retention is a determining factor for primary stability of miniscrew. Studies investigated
various ways to increase primary stability. The aim of this study is to introduce a new configuration of miniscrew
system which is believed to obtain more primary stability.
Material and Methods: Freshly ovine mandibles were cut in blocks. Twenty-seven miniscrews (diameter 1.6 × 8
mm; G2, Dual Top Anchor System, Jeil Medical, Seoul, Korea) were inserted in the blocks and divided in 2 experimental
groups: single miniscrew and the innovated design “Seifi Twin Screw (STS)”. Primary stability was
evaluated by Periotest “M”® device.
Results: Independent t-test showed a significant difference between 2 experimental groups in periotest evaluation
(p< 0.05). STS demonstrated higher primary stability due to its mechanical configuration and design.
Conclusions: The STS provides higher primary stability and was found to be effective in increased success rate of
miniscrew systems from the standpoint of primary stability.
|
|