Fracturas proximales de fémur. Osteosíntesis con tornillo-placa deslizante versus clavo gamma
Mostra el registre complet de l'element
Visualització
(353.7Kb)
|
|
|
|
|
|
Plasencia Arriba, M. A.; Juanicotena Iturralde, M.ª J.; Jativa Silvestre, F.
|
|
Aquest document és un/a article, creat/da en: 1995
|
|
|
|
We report a prospective study of ninety fractures of proximal femur treated by
either the GAMMA nail or the dynamic hip screw (DHS). We found no difference in operating
time, blood loss or stay in the hospital. However, the GAMMA nail group had a earlier full
weight-bearing, but there was no significant difference in hip function at final review. There
was failure of proximal fixation in four cases with the DHS. More intra-operative complications
were recorded in the GAMMA nail group, mainly a high incidence of femoral shaft fracture
which we relate in part to the design of the implant. We only recommend the use of the GAMMA
nail for unstable intertrochanteric fractures of the proximal femur.Presentamos un estudio retrospectivo de noventa fracturas del fémur proximal
tratados con clavo GAMMA o tornillo-placa dinámico de cadera (DHS). No encontramos diferencias
en la duración de la intervención, pérdida hemática, ni estancia hospitalaria. Sin embargo,
la serie con clavo GAMMA inició el apoyo más temprano, aunque en la revisión al final
del seguimiento no había diferencias significativas en cuanto a la función de la cadera. El fracaso
de la síntesis ocurrió en cuatro casos con el DHS, pero las complicaciones intraoperatorias
fueron más frecuentes con el clavo GAMMA, principalmente elevada incidencia de fracturas de la diáfisis femoral, asociado en parte al diseño del implante. Recomendamos el empleo del clavo
GAMMA sólo para las fracturas inestables del fémur proximal. |
|