A total of 78 high tibial osteotomies —70 for valguization and 8 for varization—
in 60 patients were reviewed. I all cases, the osteotomy was stabilized by a new staple designed
with compressive effect, due to the elasticity of the material used. The follow-up of the patients
ranged from 3 to 10 years. According to Coventry's criteria (1973), good results were obtained
in 61 osteotomies (78%), fair in 11 (14%) and poor in 6 (8%). Neither loss of correction nor
migration of the staple were detected. Complications were found as paresia of the extensor
longus of the first toe in one case, infection of the soft tissues in 2 cases and persisting
postoperative edema of the lower extremity in other 2 patients. Healing of the osteotomy was
achieved in 49 days average (range, 30-67). The new compressive staple permit to reduce or
eliminate postoperative plaster inmovilization, facilitate an early knee rehabilitation, and diminish
the risk for loss of correccion.Se revisan 60 pacientes con 78 osteotomías proximales de tibia —70 valguizantes
y 8 varizantes— estabilizadas con una grapa de nuevo diseño con propiedades compresoras,
debido a la elasticidad de su material. El seguimiento de los pacientes osciló entre los
3 y 10 años. De acuerdo con los criterios tradicionales de Coventry (1973), hemos obtenido
buenos resultados en 61 osteotomías (78%), regulares en 11 (14%) y malos en 6 (8%). No
se detectaron pérdidas de corrección, ni migraciones de la grapa. Las complicaciones fueron:
una paresia del extensor largo del primer dedo del pie, 2 infecciones superficiales de partes
blandas y 2 edemas postoperatorios persistentes de extremidad inferior. La consolidación se
obtuvo en una media de 49 días (rango 30-67). La nueva grapa compresora permite reducir
al mínimo o eliminar la inmovilización con yeso, facilita una precoz movilización postoperatoria
de la rodilla, y disminuye los riesgos de pérdida de corrección.
|