Evaluation of the validity of the Bolton Index using cone-beam computed tomography (CBCT)
Mostra el registre complet de l'element
Visualització
(872.6Kb)
|
|
|
|
|
|
Tarazona Álvarez, Beatriz; Llamas Carreras, José María; Cibrián Ortiz de Anda, Rosa María; Gandía Franco, José Luis; Paredes Gallardo, Vanessa
|
|
Aquest document és un/a article, creat/da en: 2012
|
|
|
|
Aims: To evaluate the reliability and reproducibility of calculating the Bolton Index using cone-beam computed tomography (CBCT), and to compare this with measurements obtained using the 2D Digital Method. Material and Methods: Traditional study models were obtained from 50 patients, which were then digitized in order to be able to measure them using the Digital Method. Likewise, CBCTs of those same patients were undertaken using the Dental Picasso Master 3D® and the images obtained were then analysed using the InVivoDental programme. Results: By determining the regression lines for both measurement methods, as well as the difference between both of their values, the two methods are shown to be comparable, despite the fact that the measurements analysed presented statistically significant differences. Conclusions: The three-dimensional models obtained from the CBCT are as accurate and reproducible as the digital models obtained from the plaster study casts for calculating the Bolton Index. The differences existing between both methods were clinically acceptable
|
|