Curing effectiveness of resin composites at different exposure times using LED and halogen units

Mostra el registre parcial de l'element

dc.contributor.author Ceballos García, Laura es
dc.contributor.author Fuentes Fuentes, María Victoria es
dc.contributor.author Tafalla Pastor, Héctor es
dc.contributor.author Martínez, Álvaro es
dc.contributor.author Flores, Josefa es
dc.contributor.author Rodríguez, Jesús es
dc.date.accessioned 2017-09-06T08:28:18Z
dc.date.available 2017-09-06T08:28:18Z
dc.date.issued 2009 es
dc.identifier.citation Ceballos García, Laura ; Fuentes Fuentes, María Victoria ; Tafalla Pastor, Héctor ; Martínez, Álvaro ; Flores, Josefa ; Rodríguez, Jesús. Curing effectiveness of resin composites at different exposure times using LED and halogen units. En: Medicina oral, patología oral y cirugía bucal. Ed. inglesa, 14 1 2009: 11- es
dc.identifier.uri http://hdl.handle.net/10550/60541
dc.description.abstract Objective: To compare the polymerization effectiveness of two resin composites cured with a quartz tungsten halogen (QTH) lamp or a light emitting diodes (LED) unit. Study design: Filtek Z250 (3M ESPE) and Spectrum TPH (Dentsply DeTrey) resin composites were placed in 9 mm deep and 4 mm wide metallic molds and cured using the QTH light Hilux 200 (Benlioglu) or the LED unit Smartlite IQ (Dentsply DeTrey) for 20 or 40 s (three specimens per group). Measurement of depth of cure was carried out by means of a scraping technique, according to ISO 4049. The microhardness measurements were performed using a calibrated Vickers indenter (100 g load, 30 s) at depths of 0.5, 1.5, 2.5, 3.5, 4.5 and 5.5 mm from the top of the composite in the same specimens. Results were analyzed by ANOVA, Student´s t and Student-Newman-Keuls tests (p<0.05). Results: Filtek Z250 exhibited higher depth of cure and Vickers microhardness values than Spectrum TPH under each experimental condition evaluated. Depth of cure and microhardness were not affected by the curing light used. However, hardness values were influenced by the interaction between curing light and exposure time. Specimens irradiated for 20 s exhibited higher microhardness values when the LED curing light was used. Exposure time had no influence on the microhardness values for depths from 0.5 to 2.5 mm. At higher depths, irradiation for 40 s produced greater microhardness values. Conclusions: Curing effectiveness of resin composite is not only dependent on the curing light unit. Results vary greatly with composite brand, thickness of the resin composite and the duration of the exposure. es
dc.title Curing effectiveness of resin composites at different exposure times using LED and halogen units es
dc.type journal article es_ES
dc.subject.unesco UNESCO::CIENCIAS MÉDICAS es
dc.identifier.doi es
dc.type.hasVersion VoR es_ES