NAGIOS: RODERIC FUNCIONANDO

Immediate loading and immediate restoration in 105 expanded-platform implants via the Diem System after a 16-month follow-up period

Repositori DSpace/Manakin

IMPORTANT: Aquest repositori està en una versió antiga des del 3/12/2023. La nova instal.lació está en https://roderic.uv.es/

Immediate loading and immediate restoration in 105 expanded-platform implants via the Diem System after a 16-month follow-up period

Mostra el registre parcial de l'element

dc.contributor.author Calvo Guirado, José Luis es
dc.contributor.author Ortiz Ruiz, Antonio José es
dc.contributor.author Gómez Moreno, Gerardo es
dc.contributor.author López Marí, Laura es
dc.contributor.author Bravo González, Luis Alberto es
dc.date.accessioned 2017-09-26T18:03:43Z
dc.date.available 2017-09-26T18:03:43Z
dc.date.issued 2008 es
dc.identifier.citation Calvo Guirado, José Luis ; Ortiz Ruiz, Antonio José ; Gómez Moreno, Gerardo ; López Marí, Laura ; Bravo González, Luis Alberto. Immediate loading and immediate restoration in 105 expanded-platform implants via the Diem System after a 16-month follow-up period. En: Medicina oral, patología oral y cirugía bucal. Ed. inglesa, 13 9 2008: 10- es
dc.identifier.uri http://hdl.handle.net/10550/61111
dc.description.abstract Objective: The aim of this study was to report on our experience and outcomes with Certain Prevail Implants and immediate loading via the Diem System after a 16-month follow-up period. Materials and Methods: Over a 16- month period, 105 (14 maxilla, 91 mandible) expanded-platform implants were placed in 18 patients (15 females, 3 males; 55.97 ±7.25 SD years). Resonance frecuency analysis (RFA) was measured on the day of placement and at 3, 12 and 16 months. All prostheses were screw mounted on IOL DIEM standard abutments. The follow-up time varied between 3 up to 16 months. Results: One implant (0.9%) failed during final prothetic placement (3 months). The RFA (ISQ) measurements at 16 months evalution for 4 mm- and 5 mm diameter implants were: 76.13 ± 5.0 SD and 76 ±7.7 SD respectively. The RFA data related to mandible and maxilla, the day 0 mean was 76.47 ± 4.75 SD and 70.13 ± 1.35 SD, at 3 months 70.53 ± 4.0 SD and 66.07 ±2.9 SD, at 12 months 68.23± 3.6 SD and 66.40 ±2.7 SD and 16 months 68.23± 3.6 SD and 66.60 ± 2.5 SD respectively. The crestal bone resorption was 0.6 mm at 16 months in platform switching implants.Conclusions: The cumulative implant survival rate during the follow-up period was 99.1% Immediate loading on IOL Diem abutments is a reliable and effective technique for edentulous patients in the maxilla and mandible. es
dc.title Immediate loading and immediate restoration in 105 expanded-platform implants via the Diem System after a 16-month follow-up period es
dc.type journal article es_ES
dc.subject.unesco UNESCO::CIENCIAS MÉDICAS es
dc.identifier.doi es
dc.type.hasVersion VoR es_ES

Visualització       (443.1Kb)

Aquest element apareix en la col·lecció o col·leccions següent(s)

Mostra el registre parcial de l'element

Cerca a RODERIC

Cerca avançada

Visualitza

Estadístiques