Predictability of the resonance frequency analysis in the survival of dental implants placed in the anterior non-atrophied edentulous mandible

Repositori DSpace/Manakin

IMPORTANT: Aquest repositori està en una versió antiga des del 3/12/2023. La nova instal.lació está en https://roderic.uv.es/

Predictability of the resonance frequency analysis in the survival of dental implants placed in the anterior non-atrophied edentulous mandible

Mostra el registre parcial de l'element

dc.contributor.author González García, R. es
dc.contributor.author Monje Gil, Florencio es
dc.contributor.author Moreno García, C. es
dc.date.accessioned 2017-12-16T12:10:11Z
dc.date.available 2017-12-16T12:10:11Z
dc.date.issued 2011 es
dc.identifier.citation González García, R. ; Monje Gil, Florencio ; Moreno García, C.. Predictability of the resonance frequency analysis in the survival of dental implants placed in the anterior non-atrophied edentulous mandible. En: Medicina oral, patología oral y cirugía bucal. Ed. inglesa, 16 5 2011: 6- es
dc.identifier.uri http://hdl.handle.net/10550/63682
dc.description.abstract Background: Dental primary implant stability is considered essential in the success of the osseointegration process. The recent advent of the resonance frequency analysis (RFA) seems to effectively measure primary implant stability, although its relationship with implant survival has to be further established. Patients and Methods: Seventy patients with complete mandibular edentulism underwent dental implant rehabilitation by means of the placement of 68 dental implants within the interforaminal region and subsequent placement of an overdenture. Primary implant stability was measured by means of RFA and it was expressed in terms of implant stability quotient (ISQ) on the day of the implant insertion and at the time of the healing abutment placement in a conventional implant two-stage surgical procedure. Results: Overall implant survival rate was 97.1% at the end of the follow-up period. The mean ISQ value for 3.75 and 4.25 mm diameter implants was 78.4 ± 5.46 and 80.83 ± 5.35 respectively, at the time of the implant placement; and 76.68 ± 4.34 and 78.22 ± 6.87 respectively, at the second surgical stage. No statistical differences were observed in relation to changes in mean ISQ value along the healing process Conclusions: No statistical differences in terms of primary and secondary implant stability measured by RFA exists between 3.75 mm and 4.25 mm diameter implants in the conventional implant two-stage surgical procedure in patients with non-atrophied edentulous mandible being restored with an overdenture. Furthermore, no statistical association between RFA and the implant insertion torque was observed for endosseous dental implant placement at the first surgical stage. © Medicina Oral. es
dc.title Predictability of the resonance frequency analysis in the survival of dental implants placed in the anterior non-atrophied edentulous mandible es
dc.type journal article es_ES
dc.subject.unesco UNESCO::CIENCIAS MÉDICAS es
dc.identifier.doi 10.4317/medoral.16982 es
dc.type.hasVersion VoR es_ES

Visualització       (664.6Kb)

Aquest element apareix en la col·lecció o col·leccions següent(s)

Mostra el registre parcial de l'element

Cerca a RODERIC

Cerca avançada

Visualitza

Estadístiques