NAGIOS: RODERIC FUNCIONANDO

Effect of framework type on survival probability of implant-supported temporary crowns : an in vitro study

Repositori DSpace/Manakin

IMPORTANT: Aquest repositori està en una versió antiga des del 3/12/2023. La nova instal.lació está en https://roderic.uv.es/

Effect of framework type on survival probability of implant-supported temporary crowns : an in vitro study

Mostra el registre parcial de l'element

dc.contributor.author Rodrigues, Vinicius-Aneas es
dc.contributor.author Dal Piva, Amanda-Maria-de Oliveira es
dc.contributor.author Yamaguchi, Claudio-Akira es
dc.contributor.author Borges, Alexandre- Luiz-Souto es
dc.contributor.author Mukai, Marcio-Katsuyoshi es
dc.contributor.author Tribst, João Paulo Mendes es
dc.date.accessioned 2020-10-22T11:51:13Z
dc.date.available 2020-10-22T11:51:13Z
dc.date.issued 2020 es
dc.identifier.citation Rodrigues, Vinicius-Aneas ; Dal Piva, Amanda-Maria-de Oliveira ; Yamaguchi, Claudio-Akira ; Borges, Alexandre- Luiz-Souto ; Mukai, Marcio-Katsuyoshi ; Tribst, João Paulo Mendes. Effect of framework type on survival probability of implant-supported temporary crowns : an in vitro study. En: Journal of Clinical and Experimental Dentistry, 12 5 2020: 433-439 es
dc.identifier.uri https://hdl.handle.net/10550/76051
dc.description.abstract This in vitro study evaluated the effect of framework type on the survival probability of temporary implant-supported crowns and on the implant platform structure after dynamic fatigue. Thirty (30) external hexagon implants (3.75 x 10 mm) were embedded in acrylic resin following the ISO-14801. Standardized temporary crowns (n=10, N=30) were manufactured in acrylic resin and divided according to the framework type: Total plastic, Plastic with CoCr base and Titanium. The crowns were installed onto the implants (20N.cm) and fatigued (100N, 2 Hz) to determine the crowns? survival probability for missions of 300.000 and 600.000 cycles. Fatigue data were submitted to the Kaplan-Meier test followed by Wilcoxon and Log Rank, all with ? = 5%. The implant platforms were parametrically inspected based on the scanning before and after the fatigue to evaluate the damage. The strain values were analyzed using One-way ANOVA and Tukey test, all with ? = 5%. ANOVA revealed that the Total plastic showed less implant damage (-0.07 ± -0.03 mm) than the Plastic with CoCr base (-0.08 ± -0.04 mm) and the Titanium (-0.10 ± -0.01 mm) frameworks. Therefore, the framework type to manufacture implant-supported temporary crowns influences the fatigue survival of the restoration and the implant platform damage. The Plastic with CoCr base and Titanium frameworks showed superior reliability than the Total plastic framework which could not survive 600,000 cycles. The Plastic with CoCr base and the Titanium framework are suitable for restorations over 3 months in use, without a difference in the implant platform damage. es
dc.title Effect of framework type on survival probability of implant-supported temporary crowns : an in vitro study es
dc.type journal article es_ES
dc.subject.unesco UNESCO::CIENCIAS MÉDICAS es
dc.identifier.doi 10.4317/jced.56292 es
dc.type.hasVersion VoR es_ES

Visualització       (1.068Mb)

Aquest element apareix en la col·lecció o col·leccions següent(s)

Mostra el registre parcial de l'element

Cerca a RODERIC

Cerca avançada

Visualitza

Estadístiques