NAGIOS: RODERIC FUNCIONANDO

Electronic Design for a bleeding detector to be used in intraoperative radiotherapy applications

Repositori DSpace/Manakin

IMPORTANT: Aquest repositori està en una versió antiga des del 3/12/2023. La nova instal.lació está en https://roderic.uv.es/

Electronic Design for a bleeding detector to be used in intraoperative radiotherapy applications

Mostra el registre parcial de l'element

dc.contributor.author García Gil, Rafael
dc.contributor.author Casans Berga, Silvia
dc.contributor.author Sanchis Sánchez, Enrique
dc.contributor.author Pérez Calatayud, Ignacio
dc.contributor.author Pérez Calatayud, José
dc.contributor.author Sanchis Peris, Enrique
dc.date.accessioned 2023-11-15T11:32:12Z
dc.date.available 2023-11-15T11:32:12Z
dc.date.issued 2021
dc.identifier.uri https://hdl.handle.net/10550/91164
dc.description.abstract An electronic design for a bleeding detector used with mobile linac during intraoperative radiotherapy (IORT) is presented. It consists on a readout circuit based on a De Sauty Bridge and a wireless power supply using a supercapacitor. The goal is the measurement of the height of fluid present in the applicator in order to avoid a misadministration dose during the irradiation process. A minimum measurement height of 5 mm is required by clinic to be confident with the selected radiation dose. The measurement circuit uses a capacitive sensor which detects variations lower than 0.1 pF. The readout circuit converts dynamic capacitance variation from the bleeding detector into an analog voltage signal. Tests have been performed with a 40 mm diameter polymethyl methacrylate applicator (PMMA) using water. Non-bevelled PMMA 0o and 45o tilted have been considered, being this the worst condition used in the clinic. The actual height of the fluid with respect to that measured by our system shows good linearity and repeatability within acceptable limits. For 0o tilted, a 2 mm height fluid detection is achieved well below the 5 mm height demanded by clinic. For 45o tilted, a 5 mm height fluid detection is achieved.
dc.language.iso eng
dc.relation.ispartof Ieee Sensors Journal, 2021, vol. 21, p. 4786-4792
dc.source García Gil, Rafael Casans Berga, Silvia Sanchis Sánchez, Enrique Pérez Calatayud, Ignacio Pérez Calatayud, José Sanchis Peris, Enrique (2021). Electronic Design for a bleeding detector to be used in intraoperative radiotherapy applications. Ieee Sensors Journal, 21, 4786-4792. https://doi.org/10.1109/JSEN.2020.3032828
dc.subject electrònica
dc.subject electrònica aparells i instruments
dc.subject electrònica mèdica
dc.title Electronic Design for a bleeding detector to be used in intraoperative radiotherapy applications
dc.type journal article
dc.date.updated 2023-11-15T11:32:12Z
dc.identifier.doi 10.1109/JSEN.2020.3032828
dc.identifier.idgrec 142840
dc.rights.accessRights open access

Visualització       (5.116Mb)

Aquest element apareix en la col·lecció o col·leccions següent(s)

Mostra el registre parcial de l'element