NAGIOS: RODERIC FUNCIONANDO

Evaluation of the force generated by gradual deflection of 0.016-inch NiTi and stainless steel orthodontic wires in self-ligating metallic and esthetic brackets

Repositori DSpace/Manakin

IMPORTANT: Aquest repositori està en una versió antiga des del 3/12/2023. La nova instal.lació está en https://roderic.uv.es/

Evaluation of the force generated by gradual deflection of 0.016-inch NiTi and stainless steel orthodontic wires in self-ligating metallic and esthetic brackets

Mostra el registre parcial de l'element

dc.contributor.author Francisconi, Manoela-Fávaro es
dc.contributor.author Janson, Guilherme es
dc.contributor.author Henriques, José Fernando es
dc.contributor.author Freitas, Karina-Maria-Salvatore es
dc.contributor.author Francisconi, Paulo-Afonso-Silveira es
dc.date.accessioned 2019-07-24T11:37:53Z
dc.date.available 2019-07-24T11:37:53Z
dc.date.issued 2019 es
dc.identifier.citation Francisconi, Manoela-Fávaro ; Janson, Guilherme ; Henriques, José Fernando ; Freitas, Karina-Maria-Salvatore ; Francisconi, Paulo-Afonso-Silveira. Evaluation of the force generated by gradual deflection of 0.016-inch NiTi and stainless steel orthodontic wires in self-ligating metallic and esthetic brackets. En: Journal of Clinical and Experimental Dentistry, 11 5 2019: 464-469 es
dc.identifier.uri https://hdl.handle.net/10550/71111
dc.description.abstract The purpose of this study was to evaluate the deflection forces of 0.016-inch Nitinol and stainless steel orthodontic wires, in association to different self-ligating brackets. Specimens were mounted in a clinical simulation model and evaluated in a Universal Testing Machine (INSTRON 3342), using a 10N load cell and ISO 15,841, as a protocol. Eight of these models were prepared, each one for the bonding of each set of self-ligating accessories to be tested: Damon Q, Damon Clear (Ormco), In-Ovation R, In-Ovation C (GAC), BioQuick, QuickClear (Forestadent), SmartClip and Clarity SL (3M). Data were subjected to One-way ANOVA, followed by Tukey tests (P<0.05). Elastic deflection results showed that the deactivation forces increased with increase in wire deflection in the different brackets evaluated. For the different combinations, Clarity SL generated the greatest force and Damon Clear presented the lowest force when compared to the other brackets in all alloys and deflections. BioQuick and QuicKlear were those with the most similar behavior with each other. Strength values increased with gradual increase in wire deflection in all evaluated brackets. Clarity SL generated the greatest and Damon Clear the lowest force when compared to the other brackets in all alloys and deflections tested. es
dc.title Evaluation of the force generated by gradual deflection of 0.016-inch NiTi and stainless steel orthodontic wires in self-ligating metallic and esthetic brackets es
dc.type journal article es_ES
dc.subject.unesco UNESCO::CIENCIAS MÉDICAS es
dc.identifier.doi 10.4317/jced.55698 es
dc.type.hasVersion VoR es_ES

Visualització       (1.031Mb)

Aquest element apareix en la col·lecció o col·leccions següent(s)

Mostra el registre parcial de l'element

Cerca a RODERIC

Cerca avançada

Visualitza

Estadístiques